The
High Power Frequency Response Analyzer 0355 is designed primarily
for use in AC impedance evaluation of batteries and supercapacitors,
where the accurate measurement of low impedance values is
essential. Utilizing the latest digital signal processing
techniques, and multiple algorithms developed by Maccor engineers,
it has the highest accuracy and repeatability of any Frequency
Response Analyzer instrument for this application.
The FRA 0356 is intended for use in conjunction with the Maccor
Series 4000 and Series 4300 battery test systems, to provide
a combined research tool capable of a wide range of experiments.
These include long term cycling, pulse discharging, cyclic
voltametry, differential capacity, and frequency response
analysis, working with either half or full cell samples.
The FRA 0355 is easy to set up and simple to operate. The
system operates under full computer control, and stability
is maintained with any test sample without the need to select
bandwidth, requiring no manual adjustments or settings of
dials.
- AC impedance measurements integrated with other system
experiments.
- Programmable AC signal amplitude in Peak value.
- Selectable Logarithmic Sweep.
- Programmable Range of Frequency Sweep.
- Selectable Number of Frequency Steps per Decade.
- Program time delays for the start of experiment and at
each frequency stop to allow cell to stabilize before taking
measurements through the use of rest step(s).
- Data file includes: Frequency, Time of Measurement, Magnitude
and Phase Angle, Real and Imaginary Impedance, Capacitance
and Inductance.
- View data in real time.
- Standard Plots include: Nyquist, Bode, Capacitance, Inductance,
Impedance Magnitude, Phase Angle, Real Impedance, Imaginary
Impedance; Versus: Frequency, Real Impedance, Imaginary
Impedance, Cycle, Time, Voltage, Current, Capacity, Energy,
Power.
- Frequency Range: 1mHz to 30KHz
- AC Impedance Range: 100 µOhm to 1000 Ohm
- Impedance Accuracy: ±0.1% of reading or ±0.00005%
of full scale whichever is the larger
- AC Signal Amplitude: 100 µV to 5V
- DC Voltage: -5V to 55VDC
- Peak AC Current: 3000 mA
- Peak AC Voltage 0.1mV to 5V
- Connection: 4-terminal Kelvin
If you require further
information about this product or service please contact us
here
|