The Model 4300 is a
small desk top mounted cell and materials test system. It
has been designed to provide the most economical solution
to meet the testing needs of Research Institutes,
University Research Departments, and small Test Laboratories.
Typical applications include performance testing of
electrode and separator materials, and half or full cells
of any cell chemistry. Although this system is low in cost,
it is high in features, providing most
of the capability of our larger test systems
The Model 4300 System can be configured
with 1, 2, 4 or 8 channels in a desktop unit, complete with
PC computer and software for control and data acquisition
purposes. On multi channel testers all test channels have
four (4) current ranges. This allows current set points to
be programmed from 300 nA to 5A on each test channel, on single
channel testers higher current options up to 120Amps are available.
All current ranges have an accuracy of 0.02% of full scale,
the highest accuracy available in any multi-current range
systems and 0.05% on single-current range systems
Operating independently, each test channel can be started
and stopped at any time, and can be programmed to perform
different tests from each other. Optionally, System operation
is based on Maccor's Windows-based
software, providing both sophisticated test control and
high speed data aqcuisition.
- Number of Test Channels:
1, 2, 4 or 8 independent
- Voltage Ranges:
±5 V, or 0 V to 10 V
- Voltage Accuracy:
0.02% of full scale voltage
- Voltage Resolution:
16 bit (1 part in 65,536)
- Current single-channel system
Single range: 1mA - 120A Full Scale
- Current Accuracy:
0.05% of full scale current
- Current multi-channel system
Range 1: 150 µA Full Scale ±30
nA
Range 2: 5 mA Full Scale
±1 µA
Range 3: 150 mA Full Scale ±30
µA
Range 4: 5 A Full Scale
±1 mA
- Current Accuracy:
0.02% of full scale current
- C urrent Resolution:
16 bit (1 part in 65,536)
- Time Resolution:
10 mS standard, GSM/CDMA and other pulse capabilities available
as an option
- Minimum Pulse width:
100 uS
- Rise Time:
< 30 uS
- Switch Time:
Full charge to full discharge 400 mS standard <500 uS
optional
- Data Recording Rate:
200 data points per second per system standard
- Data Recording Interval:
Time (minimum 10 mS standard, 5 mS and 1 mS available),
V, I, Ah, Wh, T, P
- Operating Modes:
Constant Current, Constant Voltage, Constant Power, Constant
Resistance, Ramp Voltage
- Number of Steps per Test:
127 standard plus subroutines
- AC Input:
110/220 VAC, 50/60 Hz, single phase
- Dimensions:
22 cm. wide x 45 cm. high x 54 cm. deep
- Weight:
Typically 15 kg
If you require further information about this product or
service please contact us here
|