Cellcare Technologies Limited
Home Company Profile Products & Services Contact Details
 
Home
Products & Services
Maccor

Programmable Multi-Channel Test Systems

 

Maccor 3600 Series
 
3600

The multifunction SERIES 3600 primary cell test system has been engineered to meet the growing demands of existing and upcoming ANSI and IEC test standards. Each test position: Operates independently of the other positions; Can be user programmed to run a wide range of multi step tests, including pulse tests; Can be user programmed to operate in steps of fixed current, fixed power, fixed voltage, fixed resistance; Has a dynamic range of over 25,000 to 1 to meet ANSI and IEC standards of current|power|resistance.

The Series 3600 is supplied as a complete turnkey system with PC computer and software. Each system is available from a desktop model to large test and quality control models with thousands of test positions. Desktop models are usually available from stock.

Powered by Maccor's standard battery testing software and utilizing similar technology as found in Maccor's high precision Series 4000 laboratory test systems. The software can be configured to automatically print a standard primary cell end of test report or to add key test data to Maccor's MIMS database. Additionally, custom data handling interfaces or user interfaces can be purchased.

The Series 3600 has a Phoenix connector interface with horizontal cell configuration providing a dense and compact configuration. Each cabinet (standard 19" rack) can have up to 192 test positions. The Series 3600 is also available in a desktop cabinet with 32-test positions.

 

 

 

System Specifications

  • Number of Test Channels:
    32 to 256 per cabinet
  • Voltage Measurement Range:
    0 V to +5 V
  • Voltage Accuracy:
    ±5 mV
  • Minimum Discharge Voltage:
    0.2 volts or 0.35 × Current, whichever is greater
  • Voltage Resolution:
    16 bit
  • Current Discharge Only
    Range 1:      1 mA Full Scale ±0.2 µA
    Range 2:     15 mA Full Scale ±3.0 µA
    Range 3:   200 mA Full Scale ±40.0 µA
    Range 4: 2500 mA Full Scale ±500 µA
  • Current Resolution:
    16 bit
  • Time Resolution:
    50 mS
  • Minimum Step Time:
    500 mS
  • Data Recording Rate:
    200 data points per second per system standard
  • Operating Modes:
    Constant Current, Constant Voltage, Constant Power, Constant Resistance
  • Number of Steps per Test:
    127 standard plus subroutines
  • Dimensions:
    22 cm. wide x 45 cm. high x 54 cm. deep

If you require further information about this product or service please contact us here

         